High-Temperature Battery Cells for Demanding Operating Conditions

High Temperatures Limit the Use of Conventional Lithium-Ion Batteries

Conventional lithium-ion batteries are stable only up to about 60 °C. Applications involving significantly higher temperatures therefore place special demands on cell chemistry, materials, and cell design.

One specific challenge lies in medical technology. Electric surgical instruments are autoclaved for sterilization and are thus exposed to high temperatures. For such applications, stability during autoclaving at 137 °C for 30 minutes is required.

High ambient temperatures can also pose a problem outside of medical technology. This includes, for example, devices used in hot regions or exposed to direct sunlight.

Temperature-Stable Cell Chemistries Enable Batteries for Thermally Demanding Applications

 

Fraunhofer ISIT is developing battery cells whose materials and design are engineered to withstand high thermal stresses. To achieve this, the electrode materials, electrolyte, separator, and cell housing are carefully matched.

The institute has developed lithium iron phosphate/lithium titanate pouch cells (LFP/LTO) with a voltage of 1.8 V, as well as lithium iron phosphate/hard carbon cylindrical cells (LFP/Hard Carbon) with a voltage of 3.2 V. Both cell types can be autoclaved multiple times at 120 °C.

For specific applications, ISIT can tailor key characteristics of the battery cell. These include, in particular, cell capacity, cell voltage, and cell geometry. This allows the design to be tailored to different technical requirements and installation scenarios.

Tailor Cell Parameters to the Application:

  • Adapt cell capacity to energy requirements
  • Design the cell voltage for the specific application
  • Adapt cell geometry to the available installation space
  • Select a suitable pouch or cylindrical cell format
  • Tailor the temperature-stable cell chemistry to the thermal requirements

 

Develop Temperature-Stable Battery Cells for Your Application

 

Optimized Cell Materials Provide Stability at High Temperatures

Temperature stability results from the interaction of the individual cell components. ISIT therefore considers the electrode materials, electrolyte, separator, and housing as part of the overall cell development process.

Established coating and battery cell manufacturing processes are used for production. ISIT's flexible cell manufacturing capabilities enable different cell concepts to be developed and key parameters to be adapted to the specific application.

Two cell concepts are currently in focus:

LFP/LTO Pouch Cells

The combination of lithium iron phosphate and lithium titanate is implemented in a pouch cell format with a cell voltage of 1.8 V. The cells can be autoclaved multiple times at 120 °C.

LFP/Hard Carbon Cylindrical Cells

The cylindrical cells combine lithium iron phosphate with hard carbon and achieve a cell voltage of 3.2 V. These cells can also be autoclaved multiple times at 120 °C.

Flexible Cell Manufacturing Enables Application-Specific Battery Cells

  • Selection of suitable electrode materials
  • Optimization of electrolyte and separator
  • Selection between pouch and cylindrical cell formats
  • Adaptation of cell voltage and capacity
  • Adaptation of cell geometry to the application

Autoclavable Battery Cells Enable Cordless Electrical Instruments in the Operating Room

A key area of application is medical technology. Temperature-stable battery cells can be used in electric surgical instruments that must be autoclaved after use. The goal is to be able to leave the battery inside the device during the sterilization process.

The “ResistantBatt” project focuses on autoclavable batteries for medical technology applications. The development is therefore geared toward electric instruments that must withstand high thermal stresses during sterilization.

In addition, temperature-stable battery cells are suitable for applications in which batteries may be exposed to temperatures above the typical operating range of conventional lithium-ion batteries. These include, among others, maritime applications and devices intended for regions with extreme heat.

 

Evaluate the Use of Temperature-Stable Battery Cells

 

Temperature-Stable Lithium-Ion Cells for Sterilizable Medical Devices

Challenge:

Battery-powered devices used in surgical environments must be sterile. Conventional lithium-ion cells can typically be used only up to approximately 60 °C and therefore cannot be sterilized together with the device in an autoclave.

Our solution:

Fraunhofer ISIT develops cell chemistries and cell designs for significantly higher temperatures. The cells are designed for temperatures above 130 °C, enabling them to withstand the thermal loads required for autoclave sterilization.

Result:

The battery cell can be sterilized along with the medical device. This simplifies handling in the operating room and can support a unified product concept across different markets.

Rechargeable High-Temperature Cells for Sensors in Drill Heads

Challenge:

Sensors in drill heads require a reliable power supply at temperatures significantly above the typical operating range of lithium-ion cells. Primary cells containing costly materials such as silver are therefore among the solutions currently used.

Our solution:

Fraunhofer ISIT develops temperature-stable cell chemistries and cell designs. The lithium-ion cells remain chemically stable at temperatures above 130 °C and are intended to replace the primary cells currently used.

Result:

Die Entwicklung schafft die Grundlage für wiederaufladbare Batteriezellen in Bohrköpfen. Dadurch können Einwegzellen sowie der Einsatz kostenintensiver Materialien reduziert werden.

The Battery Can Remain in the Device During Sterilization

Temperature-stable battery cells are designed to withstand the thermal loads of autoclaving. In appropriately designed devices, the battery therefore does not need to be removed before each sterilization cycle.

Cordless Instruments Reduce Cables in the Operating Room

Battery-powered electrical surgical instruments can operate without a permanent power cable. This can reduce the number of cables in the immediate working area.

Temperature-Stable Cell Concepts Expand the Application Range

The developed cell concepts address temperatures outside the typical operating range of conventional lithium-ion batteries. This opens up battery-powered applications for environments with high thermal loads.

Cell Voltage and Cell Geometry Can Be Tailored to the Application

ISIT can adapt cell capacity, cell voltage, and cell geometry to specific technical requirements. This enables industry partners to develop battery cells tailored to their intended application and available installation space.

Research Focus Area

Learn more about the related research focus area:

Beyond Li-B: Battery Technologies Beyond Conventional Lithium-Ion Batteries